Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Validation rules for blood smear revision after automated hematological testing using Mindray CAL-8000
Sabrina Buoro1, Tommaso Mecca1, Michela Seghezzi1
1Clinical Chemistry Laboratory, Papa Giovanni XXIII Hospital, Bergamo, Italy.
Customizing validation rules for automated blood smear review improves accuracy and workflow efficiency. This enhances hematological testing quality and optimizes laboratory processes.
Area of Science:
- Hematology
- Clinical Pathology
- Laboratory Automation
Background:
- Automated hematological testing requires robust validation rules for accurate blood smear review.
- The Mindray CAL-8000 system was evaluated for its performance in routine hematological testing.
Purpose of the Study:
- To assess the efficacy of custom-defined and International Consensus Group for Hematology (ICGH) validation rules for blood smear review.
- To analyze the workflow and performance of the Mindray CAL-8000 automated hematology system.
Main Methods:
- Analysis of 1013 peripheral blood samples using the Mindray CAL-8000 system.
- Comparison of automated results with optical microscopy (OM) findings.
- Evaluation of locally derived and ICGH validation rules.
Main Results:
- Overall agreement with locally derived rules was 91% and with ICGH rules was 85%.
- Locally derived rules showed higher sensitivity (0.97) and lower false negative rates (1.4%).
- The CAL-8000 system achieved a throughput of 208 samples/h with 14-17% requiring OM analysis.
Conclusions:
- Customization of validation rules is crucial for improving the quality of hematological testing.
- Optimizing validation rules enhances laboratory workflow efficiency and accuracy.
- The study highlights the need for tailored validation strategies in automated hematology.