Squeaking in ceramic-on-ceramic hips: No evidence of contribution from the trunnion morse taper
Lloyd Roffe1, Anthony J FitzPatrick2, Geoffrey W Rodgers2
1Department of Orthopaedic Surgery and Musculoskeletal Medicine, University of Otago Christchurch, 4345, Christchurch, New Zealand.
Abstract:
Squeaking in ceramic-on-ceramic (CoC) total hip arthroplasty (THA) was investigated with an acoustic monitoring device to distinguish between squeaking emanating from the trunnion morse taper (TMT) connection versus the articular surface. 82 patients with implant problems scheduled for revision were selected and acoustic emissions (AE) with simple movements monitored. Five of these patients with CoC implants underwent surgery and their retrieved implant components were analyzed in vitro. In vivo recordings of all 82 patients found audible squeaks produce frequency content across the entire 0-50 kHz spectrum. Of the five CoC implants the in vivo peak frequency range of flexion/extension motions was 0.8 kHz and comparable to the range of in vitro testing (0.2 kHz). In vitro TMT connection motions were very large comparatively, producing a higher average peak frequency range of 22.5 kHz. All retrieved implants showed evidence of wear at the TMT connection. These findings suggest, the TMT connection does not directly cause audible squeaking. Wear occurring at this junction may contribute to debris which disrupt lubrication, increase friction, and produce noise. This observation supports current evidence suggesting that squeaking is from the bearing surface. Our findings are the first to demonstrate that in CoC THA the recordable noise of a hip squeak does not originate nor have contribution from the TMT connection. © 2016 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 35:1793-1798, 2017.
Related Concept Videos
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Pivot Bearings
A pivot bearing is a specialized type of bearing designed to support axial loads on a rotating shaft. The bearing surface, or the pivot, is positioned at the end of a shaft to support the axial thrust. The pivot may...
Bones of the Lower Limb: Femur and Patella
Collar Bearings
The Auditory Ossicles
The aptly named stapes look very much like a stirrup. The three ossicles are unique to mammals, and each plays a role in...
Vibrating Concrete