Notes from the Field: Mycobacterium chimaera Contamination of Heater-Cooler Devices Used in Cardiac Surgery - United

Insights

Nontuberculous mycobacterial infections, including Mycobacterium chimaera, have been linked to contaminated heater-cooler devices used during open-heart surgery. These infections may result from aerosolized bacteria from the devices.

Area of Science:

  • Infectious Diseases
  • Medical Devices
  • Public Health

Background:

  • Nontuberculous mycobacterial (NTM) infections, specifically Mycobacterium chimaera, have been identified in patients undergoing open-heart surgery.
  • Clusters of invasive M. chimaera infections were reported in Switzerland and Pennsylvania, associated with heater-cooler (HC) devices.

Purpose of the Study:

  • To investigate the link between invasive NTM infections and contaminated HC devices used in open-heart surgery.
  • To alert healthcare providers and manufacturers about potential risks associated with HC devices.

Main Methods:

  • Epidemiologic and laboratory investigations were conducted to identify the source of infections.
  • Analysis of patient data and environmental samples from healthcare facilities.

Main Results:

  • A strong association was found between invasive M. chimaera infections and exposure to contaminated Stöckert 3T HC devices.
  • Contaminated HC devices, manufactured by LivaNova PLC, were identified as the likely source of M. chimaera transmission.
  • Aerosolization of M. chimaera from HC devices is the suspected route of infection, despite closed water circuits.

Conclusions:

  • Contaminated HC devices pose a significant risk for invasive M. chimaera infections in cardiac surgery patients.
  • Updated manufacturer instructions, device evaluation, and vigilance are crucial to prevent future infections.
  • Public health alerts have been issued by the FDA and CDC to address this emerging threat.