Laboratory Experience with the Liaison Analyzer in the Diagnosis of Clostridium Difficile-Associated Diarrhea
Orsolya Benedek1, Andreas Podbielski1, Philipp Warnke1
1Institute of Medical Microbiology, Virology and Hygiene, University Hospital Rostock , Rostock, Germany.
Background:
Chemiluminescent or enzyme-linked fluorescent immunoassays are commonly used to diagnose Clostridium difficile-associated diarrhea.
Methods:
The LIAISON analyzer (DiaSorin, Italy) was compared to miniVIDAS (bioMérieux, France) and, furthermore, to culture of toxigenic strains. In total, 249 native stool samples were analyzed. Sensitivities, specificities, and positive and negative predictive values were investigated. Furthermore, performance under routine conditions was assessed.
Results:
The glutamate dehydrogenase chemiluminescent immunoassay (GDH-CLIA) assay revealed a high sensitivity and negative predictive value. The toxins A&B assays exhibited approximately the same low sensitivity and high specificity. Technical drawbacks experienced with the LIAISON analyzer in 48% of the analyses considerably delayed the time to the first diagnostic report and interfered with laboratory routine workflow.
Conclusion:
The analytical performance of the investigated platforms should be reflected in the context of implementation into the laboratory workflow.