Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Clinical hemodynamic evaluation of patients implanted with a fully magnetically levitated left ventricular assist
Nir Uriel1, Sirtaz Adatya1, Jiří Malý2
1Department of Medicine, University of Chicago Medical Center, Chicago, Illinois, USA.
Background:
The HeartMate 3 (HM3) is a Conformiteé Européenne (CE) mark-approved left ventricular assist device (LVAD) with a fully magnetically levitated rotor with features consisting of a wide range of operational speeds, wide flow paths and an artificial pulse. We performed a hemodynamic and echocardiographic evaluation of patients implanted with the HM3 LVAD to assess the speed range for optimal hemodynamic support.
Methods:
Sixteen HM3 patients underwent pump speed ramp tests with right heart catheterization (including central venous pressure [CVP], pulmonary artery pressure, pulmonary capillary wedge pressure [PCWP] and blood pressure [BP]) and 3-dimensional echocardiography (3DE). Data were recorded at up to 13 speed settings. Speed changes were in steps of 100 revolutions per minute (rpm), starting at 4,600 rpm and ramping up to 6,200 rpm.
Results:
Mean original speed was 5,306 ± 148 rpm, with a majority of patients (10 of 16, 62.5%) having normal CVPs and PCWPs at their original rpm settings. Going from lowest to highest speeds, cardiac output improved at the rate of 0.08 ± 0.08 liter/min per 100 rpm (total change 1.25 ± 1.20 liters/min) and PCWP decreased at the rate of -0.48 ± 0.27 mm Hg per 100 rpm (total change -6.13 ± 3.72 mm Hg). CVP and systolic BP did not change significantly with changes in rpm. Left ventricular end-diastolic dimension (LVEDD) decreased at a rate of -0.15 ± 0.09 cm per 100 rpm. Number of rpm was adjusted based on test results to achieve CVPs and PCWPs as close to normal limits as possible, which was feasible in 13 (81.3%) patients. For the remaining 3 patients, medical management was pursued to optimize hemodynamic support.
Conclusion:
Hemodynamic normalization of pressures was achieved in the majority of patients implanted with the HM3 pump within a narrow speed range.
More Related Videos
05:18Implantation of Left Ventricular Assist Device (LVAD) in Juvenile Landrace Swine: A LVAD Implantation Model of Pediatric Heart Failure
Published on: January 16, 2026
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Cardiomyopathy II: Dilated Cardiomyopathy