Related Experiment Video
Updated: Mar 13, 2026

Author Spotlight: Noninvasive Cerebral Blood Flow Determination in Human Functional Brain Region for Diagnosis of Neurological Disorders
Published on: May 31, 2024
Improving the robustness of pseudo-continuous arterial spin labeling to off-resonance and pulsatile flow velocity
Li Zhao1, Marta Vidorreta2,3, Salil Soman1
1Department of Radiology, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, Massachusetts, USA.
Purpose:
To improve pseudo-continuous arterial spin labeling (PCASL) robustness to off-resonance and pulsatile blood flow velocity.
Methods:
Bloch equations were solved to evaluate the effect of labeling parameters in a pulsatile flow model for a range of off-resonance. Experimental confirmation was achieved in volunteers using linear phase increase between labeling pulses to approximate off-resonance errors. We first assessed the location of the labeling plane in four volunteers. Next, we explored a range of parameters-including balanced and unbalanced gradients-in five more volunteers at an optimal labeling plane location.
Results:
Simulations demonstrated that 1) high velocities are vulnerable to off-resonance, 2) unbalanced PCASL outperforms balanced PCASL, 3) increased B1 and low average gradient improve the labeling efficiency for high-velocity flow, and 4) a low ratio of selective to average gradient improves off-resonance robustness. A good agreement between theory and experiment was observed.
Conclusion:
The robustness of PCASL can be increased by selecting an unbalanced scheme with a low average gradient (0.5 mT/m), a low ratio (7×) of selective to average gradients, and the highest feasible B1 (1.8 μT). Placing the labeling plane above the carotid bifurcation and below the V3 segment, usually between the second and third vertebrae, yielded robust results. Magn Reson Med 78:1342-1351, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
Related Concept Videos
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....
Magnetic Resonance Imaging
NMR Spectrometers: Resolution and Error Correction
Measurement of Blood Pressure