Comparative performance assessment of commercially available automatic external defibrillators: A simulation and
Simone Savastano1, Vincenzo Vanni2, Roman Burkart3
1Division of Cardiology, Fondazione IRCCS Policlinico San Matteo, Pavia, Italy.
Purpose:
Early and good quality cardiopulmonary resuscitation (CPR) and the use of automated external defibrillators (AEDs) improve cardiac arrest patients' survival. However, AED peri- and post-shock/analysis pauses may reduce CPR effectiveness.
Methods:
The time performance of 12 different commercially available AEDs was tested in a manikin based scenario; then the AEDs recordings from the same tested models following the clinical use both in Pavia and Ticino were analyzed to evaluate the post-shock and post-analysis time.
Results:
None of the AEDs was able to complete the analysis and to charge the capacitors in less than 10s and the mean post-shock pause was 6.7±2.4s. For non-shockable rhythms, the mean analysis time was 10.3±2s and the mean post-analysis time was 6.2±2.2s. We analyzed 154 AED records [104 by Emergency Medical Service (EMS) rescuers; 50 by lay rescuers]. EMS rescuers were faster in resuming CPR than lay rescuers [5.3s (95%CI 5-5.7) vs 8.6s (95%CI 7.3-10).
Conclusions:
AEDs showed different performances that may reduce CPR quality mostly for those rescuers following AED instructions. Both technological improvements and better lay rescuers training might be needed.