A gate evaluation of the sources of error in quantitative 90Y PET
Jared Strydhorst1, Thomas Carlier2, Arnaud Dieudonné3
1IMIV, U1023 Inserm/CEA/Université Paris-Sud and ERL 9218 CNRS, Université Paris-Saclay, CEA/SHFJ, Orsay 91401, France.
Purpose:
Accurate reconstruction of the dose delivered by 90Y microspheres using a postembolization PET scan would permit the establishment of more accurate dose-response relationships for treatment of hepatocellular carcinoma with 90Y. However, the quality of the PET data obtained is compromised by several factors, including poor count statistics and a very high random fraction. This work uses Monte Carlo simulations to investigate what impact factors other than low count statistics have on the quantification of 90Y PET.
Methods:
PET acquisitions of two phantoms-a NEMA PET phantom and the NEMA IEC PET body phantom-containing either 90Y or 18F were simulated using gate. Simulated projections were created with subsets of the simulation data allowing the contributions of random, scatter, and LSO background to be independently evaluated. The simulated projections were reconstructed using the commercial software for the simulated scanner, and the quantitative accuracy of the reconstruction and the contrast recovery of the reconstructed images were evaluated.
Results:
The quantitative accuracy of the 90Y reconstructions were not strongly influenced by the high random fraction present in the projection data, and the activity concentration was recovered to within 5% of the known value. The contrast recovery measured for simulated 90Y data was slightly poorer than that for simulated 18F data with similar count statistics. However, the degradation was not strongly linked to any particular factor. Using a more restricted energy range to reduce the random fraction in the projections had no significant effect.
Conclusions:
Simulations of 90Y PET confirm that quantitative 90Y is achievable with the same approach as that used for 18F, and that there is likely very little margin for improvement by attempting to model aspects unique to 90Y, such as the much higher random fraction or the presence of bremsstrahlung in the singles data.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET