Kinematic analysis of diastolic function using the freely available software Echo E-waves - feasibility and
Martin G Sundqvist1, Katrin Salman2, Per Tornvall1
1Department of Clinical Science and Education, Södersjukhuset, Karolinska Institutet, and Cardiology Clinic, Södersjukhuset, SE-188 83, Stockholm, Sweden.
A new software, Echo E-waves, enables efficient analysis of early diastolic left ventricular filling using the Parameterized Diastolic Filling (PDF) method. This tool demonstrates good feasibility and accuracy for Doppler echocardiography, improving diagnostic capabilities.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Imaging
Background:
- Early diastolic left ventricular (LV) filling can be modeled using classical mechanics principles, termed the Parameterized Diastolic Filling (PDF) formalism.
- The PDF method mathematically describes E-waves from pulsed wave (PW) Doppler using LV stiffness, viscoelasticity, and load constants.
- Existing methods lacked efficient software for PDF analysis, hindering its clinical application.
Purpose of the Study:
- To develop and evaluate a freely available software application, Echo E-waves, for Parameterized Diastolic Filling (PDF) analysis.
- To assess the feasibility, time efficiency, and intra- and interobserver variability of the Echo E-waves software.
- To provide a practical tool for kinematic analysis of diastolic function using Doppler echocardiography.
Main Methods:
- Development of a semi-automatic software application (Echo E-waves) capable of analyzing DICOM files from various vendors.
- Analysis of E-waves from 20 patients by two independent investigators using the developed software.
- Assessment of analysis time, intra- and interobserver variability for derived PDF parameters.
Main Results:
- The Echo E-waves software demonstrated rapid semi-automatic analysis and results export.
- Median analysis time for a significant portion of E-waves per patient was 4.3 minutes.
- 12 out of 14 analyzed parameters showed good to excellent intra- and interobserver variability (CV 2.5-18.7%, ICC 0.80-0.99).
Conclusions:
- The implementation of the Parameterized Diastolic Filling (PDF) method in the freely available Echo E-waves software is highly feasible.
- The software offers a time-efficient solution for kinematic analysis of diastolic function.
- The Echo E-waves software exhibits good to excellent intra- and interobserver variability, supporting its clinical utility in Doppler echocardiography.
More Related Videos
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010
12:12Echocardiographic Approaches and Protocols for Comprehensive Phenotypic Characterization of Valvular Heart Disease in Mice
Published on: February 14, 2017
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...