Related Experiment Video
Updated: Mar 13, 2026

Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
Comparison of Conventional Mercury Thermometer and Continuous TherCom® Temperature Recording in Hospitalized Patients
Pradeepa H Dakappa1, Gopalkrishna K Bhat2, Ganaraja Bolumbu3
1Ph.D. Research Scholar, Department of Medicine, Kasturba Medical College, Manipal University , Mangalore, Karnataka, India .
Introduction:
Detection of accurate body temperature fluctu-ations in hospitalized patients is crucial for appropriate clinical decision-making. The accuracy and reliability of body temperature assessment may significantly affect the proper treatment.
Aim:
To compare the conventional and continuous body temperature recordings in hospitalized patients.
Materials And Methods:
This cross-sectional study was carried out at a tertiary care centre and study included 55 patients aged between 18-65 years with a history of fever admitted to a tertiary care hospital. A noninvasive continuous temperature recording was done using TherCom® device through tympanic temperature probe at tympanic site at one-minute intervals for 24 hours. The conventional temperatures were recorded in the axilla using mercury thermometer at specific time intervals at 12:00 noon, 8:00 PM and 5:00 AM. Peak temperature differences between continuous and conventional methods were compared by applying Independent sample t-test. Intra class Correlation Coefficient (ICC) test was performed to assess the reliability between two temperature-monitoring methods. A p<0.05 was considered as significant.
Results:
The average peak temperature by non-invasive continuous recording method was 39.07°C ±0.76°C while it was 37.55°C ±0.62°C by the conventional method. A significant temperature difference of 1.52°C [p<0.001;95% CI(1.26-1.78)] was observed between continuous and conventional temperature methods. Intra class Correlation Coefficient (ICC) between continuous and conventional temperature readings at 12:00 noon was α= 0.540, which had moderate reliability. The corresponding coefficients at 8:00 PM and 5:00 AM were α=0.425 and 0.435, respectively, which had poor reliability.
Conclusion:
The conventional recording of temperature is routinely practiced and does not reflect the true temperature fluctuations. However, the continuous non-invasive temperature recording is simple, inexpensive and a better tool for recording the actual temperature changes.
More Related Videos
08:22Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
Published on: October 7, 2015
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...