Differences of standard values of Supersonic shear imaging and ARFI technique - in vivo study of testicular tissue
Related Experiment Video
Updated: Mar 13, 2026

Application of Ultrasound and Shear Wave Elastography Imaging in a Rat Model of NAFLD/NASH
Published on: April 20, 2021
M Trottmann1, J Rübenthaler2, J Marcon1
1Department of Urology, Ludwig-Maximilians-University of Munich-Grosshadern Campus, Munich, Germany.
Purpose:
To investigate the difference of standard values of Supersonic shear imaging (SSI) and Acoustic Radiation Force Impulse (ARFI) technique in the evaluation of testicular tissue stiffness in vivo.
Materials And Methods:
58 healthy male testes were examined using B-mode sonography and ARFI and SSI. B-mode sonography was performed in order to scan the testis for pathologies followed by performance of real-time elastography in three predefined areas (upper pole, central portion and lower pole) using the SuperSonic® Aixplorer ultrasound device (SuperSonic Imagine, Aix-en-Provence, France). Afterwards a second assessment of the same testicular regions by elastography followed using the ARFI technique of the Siemens Acuson 2000™ ultrasound device (Siemens Health Care, Germany). Values of shear wave velocity were described in m/s. Parameters of elastography techniques were compared using paired sample t-test.
Results:
The values of SSI were all significantly higher in all measured areas compared to ARFI (p < 0.001 to p = 0.015). Quantitatively there was a higher mean SSI wave velocity value of 1,1 compared to 0.8 m/s measured by ARFI.
Conclusion:
SSI values are significantly higher than ARFI values when measuring the stiffness of testicular tissue and should only be compared with caution.