Related Experiment Video
Updated: Mar 12, 2026

A Recovery Cardiopulmonary Bypass Model Without Transfusion or Inotropic Agents in Rats
Published on: March 23, 2018
A Novel Anti-Pollution Filter for Volatile Agents During Cardiopulmonary Bypass: Preliminary Tests
Caetano Nigro Neto1, Giovanni Landoni2, Maria Angela Tardelli3
1Dante Pazzanese Institute of Cardiology, Sao Paulo, Brazil.
Objective:
Concerns regarding pollution of the operating room by volatile anesthetics and effects on atmospheric ozone depletion exist. Volatile agents commonly are used during cardiopulmonary bypass to provide anesthesia independent of any supposed myocardial protective effects. The authors' aim was to create and to assess the performance of a prototype filter for volatile agents to be connected to the cardiopulmonary bypass circuit to avoid the emission of volatile agents to the operating room, and also to the environment without causing damage to the membrane oxygenator.
Design:
Observational trial.
Setting:
University hospital.
Participants:
Prototype filter for volatile agents.
Interventions:
The prototype filter was tested in a single ex vivo experiment. The main data measured during the test were pressure drop to detect interference with the performance of the oxygenator, back pressure to detect overpressure to the outlet gas jacket of the oxygenator, analysis of exhaled sevoflurane after the membrane oxygenator, and after the filter to detect any presence of sevoflurane.
Measurements And Main Results:
The prototype filter adsorbed the sevoflurane eliminated through the outlet portion of the oxygenator. During the entire test, the back pressure remained constant (4 mmHg) and pressure drop varied from 243 mmHg to 247 mmHg.
Conclusion:
The prototype filter was considered suitable to absorb the sevoflurane, and it did not cause an overpressure to the membrane oxygenator during the test.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration
Cardiopulmonary Resuscitation II: ACLS Airway Management