Application of Biochip Microfluidic Technology to Detect Serum Allergen-specific Immunoglobulin E sIgE
Published on: April 21, 2019
Performance evaluation of ImmunoCAP® ISAC 112: a multi-site study
The ImmunoCAP® ISAC 112 assay demonstrates robust performance for measuring specific IgE (sIgE) to allergen components across multiple international sites. This multiplex immunoassay provides consistent and reproducible results, ensuring reliable allergen component diagnostics.
Area of Science:
- Allergy diagnostics
- Immunoassay technology
- Molecular allergology
Background:
- Re-introduction of the ImmunoCAP® ISAC 112 multiplex immunoassay.
- Need to evaluate its performance for specific IgE (sIgE) measurements to allergen components.
Purpose of the Study:
- To assess the performance and reproducibility of the ImmunoCAP® ISAC 112 assay across multiple international laboratory sites.
- To validate the reliability of IgE measurements to allergen components using this multiplex platform.
Main Methods:
- Multi-center study conducted at 22 European and 1 South African sites.
- Assays performed using different microarrays batches, positive/negative serum samples, and a calibration sample.
- Standardized procedures followed manufacturer's instructions for IgE measurements.
Main Results:
- Consistent results observed across all 23 sites for both positive and negative samples, with low false positive rates (0.014%).
- High reproducibility demonstrated through homogeneity analysis and clustering of identical samples.
- Good accuracy confirmed by analysis of raw continuous data, with overall study variability estimated at 25.5% after outlier filtering.
Conclusions:
- The ImmunoCAP® ISAC 112 immunoassay exhibits robust and consistent performance across diverse laboratory settings.
- Results are reliable irrespective of assay site, laboratory conditions, instruments, or microarray batches.
- The assay is suitable for accurate IgE measurements to allergen components in multi-site diagnostic applications.