Estimating the intra- and inter-individual imprecision of manual pipetting
Background:
Despite the importance of manual pipetting of fluids such as water, solutions, buffers, reagents, or biological samples in daily laboratory practice, the intra- and inter-individual imprecision of this activity has not been recently described in scientific publications.
Methods:
Twenty laboratory operators were randomly enrolled for this study. Imprecision of manual pipetting was estimated by asking each laboratory professional to dispense 1 mL, 100 μL or 10 μL of distilled water for 10 consecutive times with three certified pipettes into a 50-mL plastic container placed into a gravimetric balance. The weight of the water dispensed was systematically recorded for each of the 10 repeated attempts, and the inter- and intra-operator imprecision was finally calculated and expressed as coefficient of variation (CV%).
Results:
The mean intra-individual imprecision was 5.7% (range, 0%-11.8%) for pipetting 10 μL, 0.8% (range, 0.4%-1.9%) for pipetting 100 μL, and 0.2% (range, 0.1%-0.5%) for pipetting 1 mL. Overall, the mean inter-individual imprecision was 8.1% for pipetting 10 μL, 1.1% for pipetting 100 μL and 0.4% for pipetting 1 mL. A significantly inverse correlation was found between intra-individual pipetting imprecision and the amount of water dispensed (r = -0.80; p<0.001). No significant correlation was observed between individual pipetting performance and sex, age, qualification, and years of experience in the laboratory.
Conclusions:
The results of this study show that manual pipetting is plagued by a considerable intra- and inter-individual imprecision, which is inversely correlated with the amount of fluid dispensed.
More Related Videos
08:29An Open Source Technology Platform to Manufacture Hydrogel-Based 3D Culture Models in an Automated and Standardized Fashion
Published on: March 31, 2022
13:05Reliable Mechanochemistry: Protocols for Reproducible Outcomes of Neat and Liquid Assisted Ball-mill Grinding Experiments
Published on: January 23, 2018
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Uncertainty in Measurement: Reading Instruments
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Random and Systematic Errors
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...