Related Experiment Video
Updated: Mar 12, 2026

Author Spotlight: Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
Association Between Storage Interval and Contamination of Reprocessed Flexible Endoscopes in a Pediatric
Patricia Scanlon1, Kathleen Flaherty1, Erik A Reilly2
11Infection Prevention and Control,Boston Children's Hospital,Boston,Massachusetts.
Abstract:
OBJECTIVE The maximum safe storage interval after endoscope reprocessing remains unknown. We assessed the association between storage interval and endoscope contamination to evaluate the need for scope reprocessing prior to use. METHODS We conducted a study in 2 phases. In phase 1, we cultured 9 gastrointestinal (GI) endoscopes that had been stored for at least 7 days since reprocessing. Each scope was cultured in 3 places: external surfaces of hand piece, insertion tube, and internal channels. In phase 2, after reprocessing these scopes, we hung and cultured them prospectively in a similar fashion at 1-, 2-, 4-, 6-, and 8-week intervals without patient use. We defined clinically relevant contamination as >100 colony-forming units per milliliter (CFU/mL). RESULTS In phase 1, median hang time was 69 days (range, 8-555 days). Considering the 27 total cultures, 3 of 27 GI endoscopes (11.1%) had positive cultures, all with nonpathogenic skin flora at ≤100 CFU/mL. Median hang time was not statistically different between scopes with positive and negative cultures (P=.82). In phase 2, 7 of 131 prospective cultures (5.3%) from 6 of 9 GI endoscopes at varying storage intervals were positive, all at ≤100 CFU/mL. At 56 days after reprocessing (the longest storage interval studied), 1 of 24 cultures (4.2%) was positive (100 CFU/mL of Bacillus species from external biopsy/suction ports). CONCLUSIONS No endoscopes demonstrated clinically relevant contamination at hang times ranging from 7 to 555 days, and most scopes remained uncontaminated up to 56 days after reprocessing. Our data suggest that properly cleaned and disinfected GI endoscopes could be stored safely for longer intervals than currently recommended. Infect. Control Hosp. Epidemiol. 2017;38:131-135.
Insights
Properly reprocessed gastrointestinal endoscopes show no clinically relevant contamination even after long storage periods. This suggests current storage guidelines may be overly cautious, allowing for extended safe storage intervals.
Area of Science:
- Infection Control and Hospital Epidemiology
- Medical Device Reprocessing
- Microbiology
Background:
- The maximum safe storage duration for reprocessed endoscopes is not well-established.
- Current guidelines may necessitate frequent reprocessing, impacting workflow and resources.
Purpose of the Study:
- To assess the association between storage interval and contamination levels in reprocessed gastrointestinal (GI) endoscopes.
- To evaluate the need for reprocessing GI endoscopes immediately before use.
Main Methods:
- Two-phase study involving culturing GI endoscopes stored for varying durations (7 days to 555 days).
- Phase 1: Cultured 9 scopes stored ≥7 days. Phase 2: Prospectively cultured reprocessed scopes at 1-, 2-, 4-, 6-, and 8-week intervals.
- Clinically relevant contamination defined as >100 colony-forming units/mL (CFU/mL).
Main Results:
- In phase 1, 11.1% of scopes had low-level, nonpathogenic contamination (≤100 CFU/mL).
- In phase 2, 5.3% of cultures showed low-level contamination (≤100 CFU/mL), with only one positive culture at 56 days.
- No endoscopes demonstrated clinically relevant contamination (>100 CFU/mL) at any storage interval.
Conclusions:
- Reprocessed GI endoscopes showed no clinically relevant contamination up to 555 days of storage.
- Most scopes remained uncontaminated up to 56 days post-reprocessing.
- Findings suggest that properly cleaned and disinfected GI endoscopes can be safely stored for extended periods beyond current recommendations.
More Related Videos
09:50Induction of Intestinal Graft-versus-host Disease and Its Mini-endoscopic Assessment in Live Mice
Published on: February 11, 2019
15:49Flexible Colonoscopy in Mice to Evaluate the Severity of Colitis and Colorectal Tumors Using a Validated Endoscopic Scoring System
Published on: October 16, 2013
Related Concept Videos
Endoscopic Procedures III: Video Capsule Endoscopy
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Procedures V: ERCP
Patient...