Curtain Flow Column: Optimization of Efficiency and Sensitivity
Published on: June 12, 2016
Effect of operating parameters on a centrifugal partition chromatography separation
Norbert Fumat1, Alain Berthod1, Karine Faure1
1Univ Lyon, CNRS, Université Claude Bernard Lyon 1, Ens de Lyon, Institut des Sciences Analytiques, UMR 5280, 5 rue de la Doua, F-69100 Villeurbanne, France.
Abstract:
Centrifugal partition chromatography (CPC) is the branch of countercurrent chromatography (CCC) that works with single axis hydrostatic columns with rotary seals. The hydrodynamic of the liquid stationary phase-liquid mobile phase equilibrium in the CPC chambers has been studied theoretically and with specially designed CPC columns. In this work, we selected a simple analytical separation (no loading study) of three test solutes, coccine red, coumarin and carvone, with a commonly used heptane/ethyl acetate/methanol/water 1:1:1:1v/v biphasic liquid system and two different rotors: a commercially available 30-mL CPC instrument and a 80-mL prototype rotor designed for productivity. We fully studied this separation in many possible practical operating conditions of the two rotors, aiming at a generic column characterization. The rotor rotation was varied between 1000 and 2800rpm, the aqueous mobile phase flow rate was varied between 1 and 22mL/min with the 30-mL rotor and 10 and 55mL/min with the 80-mL rotor, the upper limits being mechanical constraints and some liquid stationary phase remaining in the rotor. The variations of Sf, the volume ratio of stationary phase in the rotor, were studied versus mobile phase flow rate and rotor rotation speed. A maximum mobile phase linear velocity was found to depend on the centrifugal field for the 30-mL rotor. This maximum velocity was not observed with the 80-mL rotor. Studying the changes in coumarin and carvone peak efficiencies, it is established that the number of cells required to make one theoretical plate, i.e. one chromatographic exchange, is minimized at maximal rotation speed and, to a lesser extent, at high mobile phase flow rate (or linear velocity). Considering the throughput, there is evidence of an optimal flow rate depending on the rotor rotation that is not necessarily the highest possible.
Related Concept Videos
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Principles Of Column Chromatography
Centrifugation
Chromatography: Introduction
The phase in which the compounds linger or on which the compounds adsorb is called the stationary phase, whereas the mobile phase is the solvent that carries the solutes to be analyzed. In traditional column chromatography, the mixture flows through the stationary phase, and the compounds partition between the stationary and mobile phases...
Analyte Adsorption and Distribution
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...