The Effect of Patient Diameter on the Dual-Energy Ratio of Selected Contrast-Producing Elements
Jack W Lambert1, Paul F FitzGerald, Peter M Edic
1From the *Department of Radiology and Biomedical Imaging, University of California, San Francisco, San Francisco, CA; and †GE Global Research, Niskayuna, NY.
Objectives:
The aim of this study was to assess whether the low- to high-kVp computed tomography (CT) number ratio at dual-energy CT is affected by changes in patient diameter.
Methods:
Seven contrast-producing elements were housed sequentially within an abdomen phantom. Fat rings enlarged the phantom diameter from 26 to 44 cm. The phantom was scanned using single-energy CT at tube potentials of 80 and 140 kVp and rapid-kVp-switching dual-energy CT.
Results:
CT numbers decreased proportionally (∼20% CT number reduction for smallest to largest phantom diameters) for low- and high-energy acquisitions but resulted in consistent dual-energy ratios for each contrast element. For 17 of 21 material pair combinations, the dual-energy ratio ranges of the two elements did not overlap, implying that discrimination should remain possible for these material pairs at all patient sizes.
Conclusions:
The dual-energy ratio for different contrast materials is largely unaffected by changes in phantom diameter. This should allow for robust separation of most contrast material combinations irrespective of patient size.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...