Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry UPLC-MS
Published on: March 14, 2013
Post-acquisition spectral stitching. An alternative approach for data processing in untargeted metabolomics by
Eirini Baira1, Eleni Siapi2, Panagiotis Zoumpoulakis2
1Department of Pharmaceutical Chemistry, Faculty of Pharmacy, School of Health Sciences, National and Kapodistrian University of Athens, Panepistimiopolis, 15771, Athens, Greece.
Introduction:
In the case of the MS-based metabolomics, the large number of false positives remains a fundamental issue.
Objective:
The aim of this study was to develop a new strategy, which highlights the number of the reliable features i.e. the detected features that correspond to a consistent peak according to chromatographic and mass spectrometric criteria.
Method:
For the analysis blood samples from 20 chickens, which were administrated with naringin and 9 samples from control, were analyzed by UHPLC-HRMS (Orbitrap Velos). Two methodologies have been compared for data processing. In the first one (classical approach), all data in the 100-900 m/z mass-to charge range were included for the data processing procedure whereas for the newly developed methodology, the data were shred in 100Da slices generating 8 datasets, which have been then subjected to the downstream MS data processing. Each dataset was treated separately and the m/z_tR features obtained by either VIP's or t-test values were merged and used as the input for the construction of the general model.
Results:
The new methodology resulted to a 4-fold increase of the peaks that could be considered chromatographically and mass spectrometrically valid.
Conclusion:
A new strategy was reported on the detection of chromatographically reliable features during a metabolomic approach. The shredding of the LC-MS chromatograms into multiple m/z ranges increased the number of the identified chromatographically reliable features.
More Related Videos
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
08:232 in 1: One-step Affinity Purification for the Parallel Analysis of Protein-Protein and Protein-Metabolite Complexes
Published on: August 6, 2018
Related Concept Videos
Tandem Mass Spectrometry
High-Resolution Mass Spectrometry (HRMS)
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...