Related Experiment Video
Updated: Mar 12, 2026

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Cardiac Implantable Electronic Device Safety during Magnetic Resonance Imaging
You Mi Hwang1, Jun Kim1, Ji Hyun Lee1
1Department of Cardiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
Background And Objectives:
Although magnetic resonance imaging (MRI) conditional cardiac implantable electronic devices (CIEDs) have become recently available, non-MRI conditional devices and the presence of epicardial and abandoned leads remain a contraindication for MRIs.
Subjects And Methods:
This was a single center retrospective study, evaluating the clinical outcomes and device parameter changes in patients with CIEDs who underwent an MRI from June 1992 to March 2015. Clinical and device related information was acquired by a thorough chart review.
Results:
A total of 40 patients, 38 with a pacemaker (including epicardially located pacemaker leads) and 2 with implantable cardioverter defibrillators, underwent 50 MRI examinations. Among the patients, 11 had MRI conditional CIEDs, while the remaining had non-MRI conditional devices. Among these patients, 23 patients had traditional contraindications for an MRI: (1) nonfunctional leads (n=1, 2.5%), (2) epicardially located leads (n=9, 22.5%), (3) scanning area in proximity to a device (n=9, 22.5%), (4) devices implanted within 6 weeks (n=2, 5%), and (5) MRI field strength at 3.0 Tesla (n=6, 15%). All patients underwent a satisfactory MRI examination with no adverse events during or after the procedure. There were no significant changes in parameters or malfunctioning devices in any patients with CIEDs.
Conclusion:
Under careful monitoring, MRI is safe to perform on patients with non-MRI conditional CIEDs, remnant leads, and epicardially located leads, as well as MRI-conditional devices.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...