Cyclic fatigue comparison among endodontic instruments with similar cross section and different surface coating
Eugenio Pedullà1, Gianluca Plotino2, Mauro Scibilia1
1Department of General Surgery and Surgical-Medical Specialties, University of Catania, Catania, Italy.
Minerva Stomatologica
|November 10, 2016
Summary
Rotary endodontic instruments Mtwo showed higher cyclic fatigue resistance than Easy Shape, with geometry being a key factor. Titanium-nitride coating did not significantly impact Easy Shape performance in cyclic fatigue tests.
Area of Science:
- Endodontics
- Dental Materials Science
- Biomaterials Engineering
Background:
- Rotary endodontic instruments are crucial for root canal preparation.
- Instrument geometry and surface treatments, like titanium-nitride (TiN) coating, can influence fatigue resistance.
- Understanding these factors is vital for preventing instrument fracture during endodontic procedures.
Purpose of the Study:
- To compare the cyclic fatigue resistance of Mtwo and Easy Shape endodontic rotary instruments.
- To analyze the influence of instrument geometry and TiN surface treatment on fatigue fracture.
- To determine if TiN coating affects the cyclic fatigue performance of Easy Shape instruments.
Main Methods:
- 130 Mtwo and Easy Shape rotary instruments of various sizes and tapers were tested for cyclic fatigue resistance.
- Time to fracture (TtF) was measured in a simulated canal (60° angle, 5 mm radius).
- Fracture surfaces were examined using scanning electron microscopy (SEM), and data analyzed with ANOVA and Tukey tests.
Main Results:
- Mtwo instruments demonstrated significantly higher cyclic fatigue resistance (higher number of cycles to fracture - NCF) compared to Easy Shape instruments of the same size and taper.
- Instruments with the same taper showed significantly higher TtF for Mtwo files than Easy Shape files.
- No significant differences in fatigue performance were found between Mtwo and Easy Shape instruments with the same tip size.
Conclusions:
- Mtwo instruments exhibit superior cyclic fatigue resistance due to their smaller core metal volume.
- Titanium-nitride (TiN) coating did not significantly influence the cyclic fatigue performance of Easy Shape instruments in this study.
- Instrument geometry, specifically core design, appears to be a more critical factor than TiN coating for cyclic fatigue resistance in these systems.