Collection optics for a Raman spectrometer based on the 90° geometry of scattered light collection
Abstract:
In the present work, efficiency of classical lens, mirror-lens, and pure mirror variants of the collection optics for a Raman spectrometer based on 90° geometry of scattered light collection is investigated. It is experimentally established that, despite a smaller collection angle, in the case of a relatively narrow input slit of the spectrometer (<100 μm), the lens optics with corrected off-axis and chromatic aberrations allows larger signal intensities to be registered. However, the low f/# mirror collection optics described in the work provide a more stable adjustment and can be used to increase the Raman signal intensities in cases when the image of the scattering volume formed by them is commensurable with the sizes of the input slit of the spectrometer.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...