Post-market surveillance to detect adverse events associated with Melody® valve implantation
Kevin D Hill1, Bryan H Goldstein2, Michael J Angtuaco3
11Department of Pediatrics, Duke University Medical Center,Durham,North Carolina,United States of America.
Objective:
The aim of this study was to describe previously unrecognised or under-recognised adverse events associated with Melody® valve implantation.
Background:
In rare diseases and conditions, it is typically not feasible to conduct large-scale safety trials before drug or device approval. Therefore, post-market surveillance mechanisms are necessary to detect rare but potentially serious adverse events.
Methods:
We reviewed the United States Food and Drug Administration's Manufacturer and User Facility Device Experience (MAUDE) database and conducted a structured literature review to evaluate adverse events associated with on- and off-label Melody® valve implantation. Adverse events were compared with those described in the prospective Investigational Device Exemption and Post-Market Approval Melody® transcatheter pulmonary valve trials.
Results:
We identified 631 adverse events associated with "on-label" Melody® valve implants and 84 adverse events associated with "off-label" implants. The most frequent "on-label" adverse events were similar to those described in the prospective trials including stent fracture (n=210) and endocarditis (n=104). Previously unrecognised or under-recognised adverse events included stent fragment embolisation (n=5), device erosion (n=4), immediate post-implant severe valvar insufficiency (n=2), and late coronary compression (n=2 cases at 5 days and 3 months after implantation). Under-recognised adverse events associated with off-label implantation included early valve failure due to insufficiency when implanted in the tricuspid position (n=7) and embolisation with percutaneous implantation in the mitral position (n=5).
Conclusion:
Post-market passive surveillance does not demonstrate a high frequency of previously unrecognised serious adverse events with "on-label" Melody® valve implantation. Further study is needed to evaluate safety of "off-label" uses.
Related Concept Videos
Pharmacovigilance
This process, termed pharmacovigilance, aims to detect, evaluate, and minimize harmful effects related to medication use. The data collection for pharmacovigilance depends on spontaneous reporting systems, where healthcare professionals or patients voluntarily report suspected ADRs.
In some cases, there...
Mitral Stenosis IV: Nursing Management
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse III: Nursing Management
Mitral Regurgitation III: Medical Management
Mitral Stenosis III: Medical Management