Impact of a motion correction algorithm on image quality in patients undergoing CT angiography: A randomized
Hussam M Sheta1, Kenneth Egstrup1, Mirza Husic1
1Department of Medical Research, OUH Svendborg Hospital, Institute of Regional Health Research, Valdemarsgade 53, 5700 Svendborg, Denmark.
Aims:
To investigate the motion correction algorithm Snapshot-Freeze (SSF) compared to standard reconstruction (STD) in patients randomized to receive beta-blockers (BB) or no beta-blockers (non-BB) before coronary CT, and to investigate if SSF can replace BB.
Methods:
One hundred and forty patients scheduled for coronary CT were randomized. All images were reconstructed by the SSF and STD algorithms. Image quality was evaluated according to Likert score (1: excellent, 2: good, 3: adequate, 4: non-diagnostic) and presence of artifacts was noted.
Results:
Images from 64 patients in the BB group (mean HR 56±4bpm) and 51 patients in the non-BB group (mean HR 67±7bpm) were analyzed. Twenty five patients were excluded because of tachycardia, bradycardia or reconstruction errors in SSF. SSF increased the number of excellent images in both groups compared to the STD algorithm (BB: 59% vs.44%; non-BB: 25% vs. 8%), but the number of non-diagnostic images was not significantly reduced. SSF reduced motion artifacts (BB: 11% vs. 31%; non-BB: 49% vs. 75%), but despite this reduction, motion artifacts in non-BB were still more frequent compared to the BB group analyzed by STD (49% vs. 31%).
Conclusion:
SSF improves image quality and reduces motion artifacts, but does not compensate for the absence of BB.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...