Effect of verification imaging on in vivo dosimetry results using Gafchromic EBT3 film
Trent Aland1, Rebecca Moylan2, Tanya Kairn3
1Icon Integrated Cancer Care, 9 MeLennan Court, North Lakes, Qld 4509, Australia.; School of Chemistry, Physics, and Mechanical Engineering, Queensland University of Technology, GPO Box 2434, Brisbane, Qld 4000, Australia; Epworth Radiation Oncology, 32 Erin Street, Richmond, Vic 3121, Australia.
Abstract:
In this work, the apparent treatment dose that kV planar or CBCT imaging contributes to Gafchromic EBT3 film used for in vivo dosimetry, was investigated. Gafchromic EBT3 film pieces were attached to a variety of phantoms and irradiated using the linear accelerator's built-in kV imaging system, in both kV planar mode and CBCT mode. To evaluate the sensitivity of the film in the clinical scenario where dose contributions are received from both imaging and treatment, additional pieces of film were irradiated using base doses of 50cGy and then irradiated using selected kV planar and CBCT techniques. For kV planar imaging, apparent treatment doses of up to 3.4cGy per image pair were seen. For CBCT, apparent treatment doses ranged from 0.22cGy to 3.78cGy. These apparent doses were reproducible with and without the inclusion of the 50cGy base dose. The contribution of apparent treatment dose from both planar kV as well as CBCT imaging can be detected, even in conjunction with an actual treatment dose. The magnitude of the apparent dose was found to be dependent on patient geometry, scanning protocol, and measurement location. It was found that the apparent treatment dose from the imaging could add up to 8% of additional uncertainty to the in vivo dosimetry result, if not taken into account. It is possible for this apparent treatment dose to be accounted for by subtraction of the experimentally determined apparent doses from in vivo measurements, as demonstrated in this work.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Biological Effects of Radiation
Imaging Studies III: Computed Tomography