Comparison of pneumotachography and anemometery for flow measurement during mechanical ventilation with volatile anesthetics
Related Experiment Video
Updated: Mar 12, 2026

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
Jarred R Mondoñedo1, Jacob Herrmann2, John S McNeil3
1Department of Biomedical Engineering and School of Medicine, Boston University, 44 Cummington Mall, Boston, MA, 02215, USA.
Abstract:
Volatile anesthetics alter the physical properties of inhaled gases, such as density and viscosity. We hypothesized that the use of these agents during mechanical ventilation would yield systematic biases in estimates of flow ([Formula: see text]) and tidal volume (V T) for two commonly used flowmeters: the pneumotachograph (PNT), which measures a differential pressure across a calibrated resistive element, and the hot-wire anemometer (HWA), which operates based on convective heat transfer from a current-carrying wire to a flowing gas. We measured [Formula: see text] during ventilation of a spring-loaded mechanical test lung, using both the PNT and HWA placed in series at the airway opening. Delivered V T was estimated from the numerically-integrated [Formula: see text]. Measurements were acquired under baseline conditions with room air, and during ventilation with increasing concentrations of isoflurane, sevoflurane, and desflurane. We also evaluated a simple compensation technique for HWA flow, which accounted for changes in gas mixture density. We found that discrepancies in estimated V T between the PNT and HWA occurred during ventilation with isoflurane (6.3 ± 3.0%), sevoflurane (10.0 ± 7.3%), and desflurane (25.8 ± 17.2%) compared to baseline conditions. The magnitude of these discrepancies increased with anesthetic concentration. A simple compensation factor based on density reduced observed differences between the flowmeters, regardless of the anesthetic or concentration. These data indicate that the choice and concentration of anesthetic agents are primary factors for differences in estimated V T between the PNT and HWA. Such discrepancies may be compensated by accounting for alterations in gas density.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Pipe Flowrate Measurement
The orifice meter is a simple,...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...