Impact of iterative reconstructions on objective and subjective emphysema assessment with computed tomography: a
Steve P Martin1, Joanna Gariani2, Anne-Lise Hachulla2
1Division of Radiology, Department of Imaging and Medical Information Sciences, Geneva University Hospitals, Geneva, Switzerland. Steve.Martin@hcuge.ch.
Objectives:
To prospectively evaluate the impact of iterative reconstruction (IR) algorithms on pulmonary emphysema assessment as compared to filtered back projection (FBP).
Methods:
One hundred ten unenhanced chest CT examinations were obtained on two different scanners. Image reconstructions from a single acquisition were done with different levels of IR and compared with FBP on the basis of the emphysema index (EI), lung volume and voxel densities. Objective emphysema assessment was performed with 3D software provided by each manufacturer. Subjective assessment of emphysema was performed as a blinded evaluation. Quantitative and subjective values were compared using repeated ANOVA analysis, Bland-Altman analysis and Kendall's coefficient of concordance (W).
Results:
Lung volumes are stable on both units, throughout all IR levels (P ≥ 0.057). EI significantly decreases on both units with the use of any level of IR (P < 0.001). The highest levels of IR are responsible for a decrease of 33-36 % of EI. Significant differences in minimal lung density are found between the different algorithms (P < 0.003). Intra- and inter-reader concordance for emphysema characterisation is generally good (W ≥ 0.77 and W ≥ 0.86, respectively).
Conclusions:
Both commercially available IR algorithms used in this study significantly changed EI but did not alter visual assessment compared to standard FBP reconstruction at identical radiation exposure.
Key Points:
• Objective quantification of pulmonary emphysema is sensitive to iterative reconstructions • Subjective evaluation of pulmonary emphysema is not influenced by iterative reconstructions • Consistency in reconstruction algorithms is of paramount importance for pulmonary emphysema monitoring.
More Related Videos
08:05Lung CT Segmentation to Identify Consolidations and Ground Glass Areas for Quantitative Assesment of SARS-CoV Pneumonia
Published on: December 19, 2020
10:37Automated Measurement of Pulmonary Emphysema and Small Airway Remodeling in Cigarette Smoke-exposed Mice
Published on: January 16, 2015
Related Concept Videos
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...