Study Design:
Biomechanical analysis.
Objectives:
To show the role of additional rods and long-term fatigue strength to prevent the instrumentation failure on three-column osteotomies.
Summary Of Background Data:
Three-column osteotomy such as pedicle subtraction osteotomy (PSO) and vertebral column resections are surgical correction options for fixed spinal deformity. Posterior fixation for the PSO involves pedicle screw-and rod-based instrumentation, with the rods being contoured to accommodate the accentuated lordosis. Pseudarthrosis and instrumentation failure are known complications of PSO.
Methods:
Unilateral pedicle screw and rod constructs were mounted in ultra-high-molecular-weight polyethylene blocks using a vertebrectomy model with the rods contoured to simulate posterior fixation of a PSO. Each construct was cycled under a 200 N load at 5 Hz in simulated flexion and extension to rod failure. Three configurations (n = 5) of titanium alloy rods were tested: single rod (control), double rod, and bridging rod. Outcomes were total cycles to failure and location of rod failure.
Results:
Double-rod and bridging-rod constructs had a significantly higher number of cycles to failure compared with the single-rod construct (p < .05). Single-rod constructs failed at or near the rod bend apex, whereas the majority of double-rod and bridging-rod constructs failed at the screw-rod or rod-connector junction.
Conclusions:
Double-rod and bridging-rod constructs are more resistant to fatigue failure compared with single-rod constructs in PSO instrumentation and could be considered to mitigate the risk of instrumentation failure.