Development and Validation of a Quantitative PCR Method for Equid Herpesvirus-2 Diagnostics in Respiratory Fluids
Published on: March 17, 2016
29.2K
Progress in Quantitative Viral Load Testing: Variability and Impact of the WHO Quantitative International Standards
R T Hayden1, Y Sun2, L Tang2
1Department of Pathology, St. Jude Children's Research Hospital, Memphis, Tennessee, USA Randall.Hayden@stjude.org.
Journal of Clinical Microbiology
|November 18, 2016
Summary
WHO quantitative standards aim to improve viral load testing accuracy. However, significant interlaboratory variability persists across multiple viruses, though some improvement is noted with international unit calibration.
Area of Science:
- Virology
- Molecular Diagnostics
- Clinical Laboratory Science
Background:
- Interlaboratory agreement in viral load testing is crucial for patient management.
- The impact of World Health Organization (WHO) quantitative standards on this agreement remains under-evaluated.
- Quantitative PCR (qPCR) assays are widely used for viral load quantification.
Purpose of the Study:
- To assess interlaboratory variability in quantitative PCR assays for common viruses.
- To evaluate the effect of WHO quantitative standards and international unit calibration on result variability.
- To identify challenges in viral load testing accuracy.
Main Methods:
- Analysis of proficiency testing survey data from 554 laboratories.
- Evaluation of quantitative PCR assay results for cytomegalovirus (CMV), Epstein-Barr virus (EBV), BK virus (BKV), adenovirus (ADV), and human herpesvirus 6 (HHV6).
- Stratification of results by assay manufacturer and calibration to international units/ml.
Main Results:
- High interlaboratory variability was observed for all tested viruses, with ranges up to 5.66 log10 copies/ml.
- Adoption of international units/ml showed some improvement in result variability, particularly for EBV.
- Variability remains a significant challenge across all tested viral load assays.
Conclusions:
- Interlaboratory variability in viral load testing is a persistent issue despite the availability of WHO quantitative standards.
- Calibration to international units/ml can help reduce variability, with varying degrees of impact across different viruses.
- Further efforts are needed to enhance standardization and improve the reliability of viral load testing.