Related Experiment Video
Updated: Mar 11, 2026

The Stroke Preclinical Assessment Network Multi-laboratory Model of Thromboembolic Stroke with Thrombolysis: TE-MCAo
Published on: December 19, 2025
Techniques for transcatheter retrieval of the occlutech ASD device United Kingdom-European multicenter report
Suhair O Shebani1, Rizwan Rehman1, Demetris Taliotis2
1Department of Congenital Heart Disease, Glenfield Hospital, University Hospitals of Leicester, Leicester, United Kingdom.
Aims:
To gather current experience in Occlutech ASD device retrieval, to determine whether snaring is an effective technique and to highlight alternative retrieval techniques; METHODS AND RESULTS: United Kingdom and European Occlutech ASD device implanters reported their experience in dealing with device embolization and retrieval. Six operators reported 12 retrieval cases. Retrieval was successful in 92% (11/12), although in most cases it was not straightforward and required multiple attempts using different techniques and equipment. When each different technique or equipment combination was considered separately, there were a total of 23 retrieval attempts. Fifteen attempts involved snaring the ball on the right atrial disc of the device ("the RA pin"). In 12/15 of these attempts the snare slipped off the RA pin. In 8/15 attempts snaring eventually failed. In two cases retrieval was facilitated by elongating the device in a blood vessel. In three cases retrieval was achieved by grasping the RA pin with the jaws of the Occlutech Flex II delivery cable; CONCLUSIONS: Snares do not grip the RA pin sufficiently to reliably retrieve the device. Funnelling the device into a blood vessel or grasping the RA pin with the jaws of the delivery cable may be successful alternatives. © 2016 Wiley Periodicals, Inc.

