Maintaining Accurate Long-Term Sensing Ability Despite Significant Size Reduction of Implantable Cardiac Monitors
Julia Lortz1, Varnavas Varnavas1, Wilko WEIßENBERGER2
1Department of Cardiology, West-German Heart and Vascular Center, University Hospital Essen, Essen, Germany.
Background:
The Reveal LINQ™ implantable cardiac monitor (ICM; Medtronic, Minneapolis, MN, USA) is obviously smaller than its precursor (the Reveal XT™), but little is known about its long-term safety. Here, we investigated the long-term R-wave sensing reliability of the Reveal LINQ™ ICM.
Methods:
We analyzed the sensing quality of the Reveal LINQ™ ICM over time between March 2014 and January 2015.
Results:
A total of 30 patients underwent Reveal LINQ™ implantation. The main reason for implantation was unexplained syncope (60%). We evaluated a total of 305.8 patient-months (median, 10.7 months) of R-wave sensing quality. A comparison of baseline and follow-up R-wave sensing values revealed no difference (0.401 mV ± 0.244 mV vs 0.447 mV ± 0.323 mV; P = 0.225).
Conclusions:
The consistent detection of events is an important safety feature of an ICM and linked to secure R-wave sensing. Despite the noticeable size reduction, the Reveal LINQ™ ICM retains reliable sensing quality over time.
More Related Videos
08:42The DREAM Implant: A Lightweight, Modular, and Cost-Effective Implant System for Chronic Electrophysiology in Head-Fixed and Freely Behaving Mice
Published on: July 26, 2024
08:43Long-term Continuous EEG Monitoring in Small Rodent Models of Human Disease Using the Epoch Wireless Transmitter System
Published on: July 21, 2015
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...