Background:
Airway exchange catheters (AEC) and fiberoptic bronchoscopes (FOB) for tracheal intubation are selected so that there is only a minimal gap between their outer and inner diameter of endotracheal tube (ETT) to minimize the risk of impingement during airway instrumentation. This study aimed to test the ease of passage of FOBs and AECs through paediatric ETT of different sizes and from different manufacturers when using current recommendations for dimensional equipment compatibility taken from text books and manufacturers information.
Methods:
Twelve different brands of cuffed and uncuffed ETT sized ID 2.5 to 5.0 mm were evaluated in an in vitro set-up. Ease of device passage as well as the locations of an impaired passage within the ETT were assessed. Redundant samples were used for same sized ETT and all measurements were triple-checked in randomized order.
Results:
In total, 51 paired samples of uncuffed as well as cuffed paediatric ETT were tested. There were substantial differences in the ease of ETT passage concordantly for FOBs and AECs among different manufacturers, but also among the product lines from the same manufacturer for a given ID size. Restriction to passage most frequently was found near the endotracheal tube tip or as a gradually increasing resistance along the ETT shaft.
Conclusions:
Current recommendations for dimensional equipment compatibility AECs and FOBs with ETTs do not appear to be completely accurate for all ETT brands available. We recommend that specific equipment combinations always must be tested carefully together before attempting to use them in a patient.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Tracheostomy Suctioning II: Procedure