Pre-Launch Absolute Calibration of CCD/CBERS-2B Sensor
Flávio Jorge Ponzoni1, Bráulio Fonseca Carneiro Albuquerque2
1Instituto Nacional de Pesquisas Espaciais, Divisão de Sensoriamento Remoto, Avenida dos Astronautas 1758, 12227-010, São José dos Campos, SP, Brazil. flavio@dsr.inpe.br.
Pre-launch absolute calibration coefficients for the Charge-Coupled Device (CCD)/China-Brazil Earth Resources Satellite 2B (CBERS-2B) sensor were determined using radiometric measurements. This marks the first calculation using Brazilian image processing criteria, enabling future data comparison and monitoring.
Area of Science:
- Remote Sensing
- Satellite Calibration
- Radiometry
Background:
- Accurate calibration of Earth observation sensors is crucial for reliable data.
- The CBERS-2B satellite carries a CCD sensor for Earth imaging.
- Pre-launch calibration provides baseline performance data.
Purpose of the Study:
- To calculate pre-launch absolute calibration coefficients for the CCD/CBERS-2B sensor.
- To implement Brazilian image processing criteria for sensor calibration.
- To establish a baseline for future vicarious calibration and data monitoring.
Main Methods:
- Radiometric measurements were conducted in a satellite integration and test hall.
- An illuminated integrating sphere was used to capture calibration imagery.
- A relative calibration procedure was applied to equalize detector responses.
- Digital numbers (DN) were correlated with radiance levels to derive absolute coefficients.
Main Results:
- Pre-launch absolute calibration coefficients for the CCD/CBERS-2B sensor were successfully calculated.
- The calibration incorporated Brazilian image processing standards for the first time.
- Established baseline coefficients for comparison with future vicarious calibration results.
Conclusions:
- The calculated pre-launch coefficients provide a critical reference for CCD/CBERS-2B data quality.
- The application of Brazilian criteria ensures consistency with regional processing standards.
- Future comparisons will facilitate sensor performance monitoring and timely data updates for users.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Uncertainty in Measurement: Reading Instruments
Accuracy and Precision
Calibration Curves: Correlation Coefficient
Common Leveling Mistakes and Errors