Related Experiment Video
Updated: Mar 11, 2026

Brachial Artery Catheterization in Swine
Published on: March 30, 2019
External and internal compressions to retrieve a kinked catheter in the right brachial artery
Keisuke Nakabayashi1, Naomi Kawakatsu2, Naomi Nakazawa2
1Department of Cardiology, Seirei Hamamatsu General Hospital, 2-12-12 Sumiyoshi, Naka-ku, Hamamatsu City, Shizuoka, 430-8558, Japan. keisuke2018@gmail.com.
Abstract:
Kinking of the catheter due to excessive rotation is not a rare complication. However, percutaneous retrieval of a kinked catheter can be difficult. The key to bailout is fixation of the catheter tip. Herein, we present a 78-year-old woman who had this complication during transradial angiography. Retrieval using several previously reported techniques was unsuccessful. We finally retrieved the kinked catheter by fixing the tip of the catheter, using external and internal compressions. The former comprises manual compression on the axillary artery, while the latter comprises deployment of a balloon catheter via another puncture site.
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization IV: Nursing Management
Cardiac Catheterization III: Left Heart Catheterization
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Cardiac Catheterization I: Pre-Procedure Overview
Assessment of blood pressure in brachial artery(two-step method)