Hypertensive Crisis During Norepinephrine Syringe Exchange: A Case Report
Roland A Snijder1, Johannes T A Knape, Toine C G Egberts
1From the *Department of Medical Technology and Clinical Physics; †Department of Anaesthesiology, Intensive Care and Emergency Medicine; ‡Department of Clinical Pharmacy, University Medical Center Utrecht, Utrecht, the Netherlands; and §Division of Pharmacoepidemiology and Clinical Pharmacology, Utrecht Institute for Pharmaceutical Sciences, Faculty of Science, Utrecht University, Utrecht, the Netherlands.
A norepinephrine overdose caused a hypertensive crisis in a patient due to a syringe pump error. In vitro tests revealed significant dosing errors, especially without an intravenous clamp, highlighting the need for vigilance.
Area of Science:
- Critical care medicine
- Biomedical engineering
- Clinical pharmacology
Background:
- Syringe pumps are crucial for precise medication delivery in critical care.
- Medication errors, including overdoses, can have severe patient consequences.
- Norepinephrine is a potent vasopressor used to manage life-threatening hypotension.
Observation:
- A patient experienced a hypertensive crisis due to a norepinephrine overdose from a syringe pump malfunction.
- The malfunction occurred during a syringe exchange when the pump repeatedly reported an error.
- An object interfering with the plunger-syringe driver interface was hypothesized as the cause.
Findings:
- In vitro testing confirmed that an object can cause significant peak dosing errors (up to +572%).
- Using an intravenous line clamp resulted in a moderate overdose (0.07 mL, +225%).
- Without a clamp, a large overdose (0.8 mL, +2700%) occurred, demonstrating the critical role of clamping.
Implications:
- Clinicians must be aware of potential syringe pump errors during medication administration.
- Implementing a clamp on the intravenous line during syringe changes can mitigate overdose risks.
- Further investigation into syringe pump error mechanisms and prevention strategies is warranted.
More Related Videos
07:59Acetylcholine Re-Challenge After Intracoronary Nitroglycerine Administration
Published on: April 4, 2022
11:08Novel Approach for Simultaneous Recording of Renal Sympathetic Nerve Activity and Blood Pressure with Intravenous Infusion in Conscious, Unrestrained Mice.
Published on: February 14, 2018
Related Concept Videos
Adrenergic Agonists: Therapeutic Uses
Emergency and Intensive Care Unit (ICU) applications: Pressor agents increase blood pressure, heart rate, and contractility in shock and organ failure situations. Dopamine can induce vasodilation and stimulate adrenoceptors. Endogenous catecholamines are effective in treating cardiogenic shock. α2-agonists like clonidine can reverse anesthesia-induced hypertension.
Allergies and...
Hypertension I: Introduction
Drugs Affecting Neurotransmitter Release or Uptake
Acute Coronary Syndrome V: Nursing Management
Adrenergic Neurons: Neurotransmission
Synthesis: Catecholamine synthesis requires tyrosine, which...
Hypertension V: Nursing Management