Evaluation of the Effect of Tumor Position on Standardized Uptake Value Using Time-of-Flight Reconstruction and Point
Yasuharu Wakabayashi1, Kenichi Kashikura2, Yasuyuki Takahashi2
1Division of Radiological Technology, Saitama Prefectural Cancer Center, Saitama, Japan; Graduate School of Radiological Technology, Gunma Prefectural College of Health Sciences, Gunma, Japan.
Objectives:
The present study was conducted to examine whether the standardized uptake value (SUV) may be affected by the spatial position of a lesion in the radial direction on positron emission tomography (PET) images, obtained via two methods based on time-of-flight (TOF) reconstruction and point spread function (PSF).
Methods:
A cylinder phantom with the sphere (30 mm diameter), located in the center was used in this study. Fluorine-18 fluorodeoxyglucose (18F-FDG) concentrations of 5.3 kBq/ml and 21.2 kBq/ml were used for the background in the cylinder phantom and the central sphere respectively. By the use of TOF and PSF, SUV and SUV were determined while moving the phantom in a horizontal direction (X direction) from the center of field of view (FOV: 0 mm) at 50, 100, 150 and 200 mm positions, respectively. Furthermore, we examined 41 patients (23 male, 18 female, mean age: 68±11.2 years) with lymph node tumors, who had undergone 18F-FDG PET examinations. The distance of each lymph node from FOV center was measured, based on the clinical images.
Results:
As the distance of a lesion from the FOV center exceeded 100 mm, the value of SUV , which was obtained with the cylinder phantom, was overestimated, while SUV by TOF and/or PSF was underestimated. Based on the clinical examinations, the average volume of interest was 8.5 cm3. Concomitant use of PSF increased SUV and SUV by 27.9% and 2.8%, respectively. However, size of VOI and distance from the FOV center did not affect SUV or SUV in clinical examinations.
Conclusion:
The reliability of SUV quantification by TOF and/or PSF decreased, when the tumor was located at a 100 mm distance (or farther) from the center of FOV. In clinical examinations, if the lymph node was located within 100 mm distance from the center of FOV, SUV remained stable within a constantly increasing range by use of both TOF and PSF. We conclude that, use of both TOF and PSF may be helpful.
More Related Videos
10:04Radionuclide-fluorescence Reporter Gene Imaging to Track Tumor Progression in Rodent Tumor Models
Published on: March 13, 2018
09:03Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET