Characterizing Far-infrared Laser Emissions and the Measurement of Their Frequencies
Published on: December 18, 2015
Detection of gaseous elemental mercury using a frequency-doubled green diode laser
Abstract:
We demonstrate a second-harmonic-generation (SHG) based method for the detection of gaseous elemental mercury by using a newly available green diode laser. Multimode ultraviolet radiation at 253.7 nm is generated through a process of SHG. Correlation spectroscopy is introduced into the scheme to guarantee the measurement accuracy. The limit of detection achieved is 0.6 μg/m3 (0.07 ppb) for 1-m pathlength and 10-s integration time. The measurement accuracy is estimated to be 1.2%. The linear response range is estimated to be 0~60 μg/m2 (6.7 ppb·m), within which the linearity error is less than 1%. Real-time monitoring of mercury volatilization is demonstrated with a time resolution of 1 s. The results of performance characterization show that the proposed method has great potentials for mercury sensing in environmental and industrial fields.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...