Technical Note: Magnetic field effects on Gafchromic-film response in MR-IGRT
Francisco J Reynoso1, Austen Curcuru1, Olga Green1
1Department of Radiation Oncology, Washington University School of Medicine, St. Louis, Missouri 63110.
Purpose:
Magnetokinetic changes may affect crystal orientation and polymerization within the active layer of radiochromic film (RCF). This effect is investigated in a magnetic resonance image-guided radiotherapy unit within the context of film dosimetry.
Methods:
Gafchromic EBT2 RCF was irradiated in a 30 × 30 × 30 cm3 solid water phantom using a Co-60 MRI guided radiotherapy system (B = 0.35 T) under normal operating conditions, and under the exact conditions and setup without a magnetic field. Fifteen 20.3 × 25.4 cm2 EBT2 film sheets were placed at three different depths (d = 0.5, 5, and 10 cm) using five different treatment plans. The plans were computed using the MRIdian (ViewRay, Inc.) treatment planning system to deliver doses between 0 and 17.6 Gy. Films were analyzed before and after irradiation to obtain the net optical density (netOD) for each color channel separately. Scanning electron microscope (SEM) images were obtained to compare the active layer of selected samples.
Results:
The results indicated that the red channel netOD decreased between 0.013 and 0.123 (average of 0.060 ± 0.033) for doses above 2.8 Gy, with a linear increase in this effect for higher doses. Green channel netOD showed similar results with a decrease between 0.012 and 0.105 (average of 0.041 ± 0.027) for doses above 3.5 Gy. The blue channel showed the weakest effect with a netOD decrease between 0.013 and 0.029 (average of 0.020 ± 0.006) for doses above 8.0 Gy. SEM images show changes in crystal orientation within active layer in RCF exposed in a magnetic field.
Conclusions:
The presence of a magnetic field affects crystal orientation and polymerization during irradiation, where netOD decreased by an average of 8.7%, 8.0%, and 4.3% in the red, green, and blue channels, respectively. The under response was dependent on dose and differed by up to 15% at 17.6 Gy.
More Related Videos
11:27Preparation and In Vitro Characterization of Dendrimer-based Contrast Agents for Magnetic Resonance Imaging
Published on: December 4, 2016
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Atomic Nuclei: Magnetic Resonance
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...
Imaging Studies IV: Magnetic Resonance Imaging