Test-retest reliability of tibiofemoral joint space width measurements made using a low-dose standing CT scanner
Neil A Segal1,2, John Bergin3, Andrew Kern3
1Department of Rehabilitation Medicine, University of Kansas Medical Center, 3901 Rainbow Boulevard, Mailstop 1046, Kansas City, KS, 66160, USA. segal-research@kumc.edu.
Objective:
To determine the test-retest reliability of knee joint space width (JSW) measurements made using standing CT (SCT) imaging.
Subjects And Methods:
This prospective two-visit study included 50 knees from 30 subjects (66% female; mean ± SD age 58.2 ± 11.3 years; BMI 29.1 ± 5.6 kg/m2; 38% KL grade 0-1). Tibiofemoral geometry was obtained from bilateral, approximately 20° fixed-flexed SCT images acquired at visits 2 weeks apart. For each compartment, the total joint area was defined as the area with a JSW <10 mm. The summary measurements of interest were the percentage of the total joint area with a JSW less than 0.5-mm thresholds between 2.0 and 5.0 mm in each tibiofemoral compartment. Test-retest reliability of the summary JSW measurements was assessed by intraclass correlation coefficients (ICC 2,1) for the percentage area engaged at each threshold of JSW and root-mean-square errors (RMSE) were calculated to assess reproducibility.
Results:
The ICCs were excellent for each threshold assessed, ranging from 0.95 to 0.97 for the lateral and 0.90 to 0.97 for the medial compartment. RMSE ranged from 1.1 to 7.2% for the lateral and from 3.1 to 9.1% for the medial compartment, with better reproducibility at smaller JSW thresholds.
Conclusion:
The knee joint positioning protocol used demonstrated high day-to-day reliability for SCT 3D tibiofemoral JSW summary measurements repeated 2 weeks apart. Low-dose SCT provides a great deal of information about the joint while maintaining high reliability, making it a suitable alternative to plain radiographs for evaluating JSW in people with knee OA.