Improving the off-axis spatial resolution and dynamic range of the NIF X-ray streak cameras (invited)
A G MacPhee1, A K L Dymoke-Bradshaw2, J D Hares2
1Lawrence Livermore National Laboratory, P.O. Box 808, Livermore, California 94551-0808, USA.
Abstract:
We report simulations and experiments that demonstrate an increase in spatial resolution of the NIF core diagnostic x-ray streak cameras by at least a factor of two, especially off axis. A design was achieved by using a corrector electron optic to flatten the field curvature at the detector plane and corroborated by measurement. In addition, particle in cell simulations were performed to identify the regions in the streak camera that contribute the most to space charge blurring. These simulations provide a tool for convolving synthetic pre-shot spectra with the instrument function so signal levels can be set to maximize dynamic range for the relevant part of the streak record.