Operation of bolometer system using Pt foil on SiN substrate detector for EAST tokamak
Related Experiment Video
Updated: Oct 6, 2026

Applying X-ray Imaging Crystal Spectroscopy for Use as a High Temperature Plasma Diagnostic
Published on: August 25, 2016
1Institute of Plasma Physics, Chinese Academy of Sciences, P.O. Box 1126, Hefei 230031, China.
Abstract:
The foil resistive bolometer diagnostic on experimental advanced superconducting tokamak has been upgraded partly with a new generation of detectors. The new detectors have faster response time. However, the microwave interference is still a serious issue for the bolometer system. The system response to microwave is tested, and the test results show that the closed Wheatstone bridge circuit in the detector is the most sensitive component to high power microwave field. Simulation results of microwave transmission by the high frequency structure simulator software and shielding design are also presented.