Related Experiment Video
Updated: Aug 16, 2026

Meso-Scale Particle Image Velocimetry Studies of Neurovascular Flows In Vitro
Published on: December 3, 2018
[A dynamic phantom for quality control of ventriculography examinations]
Abstract:
The authors designed and constructed device simulating the activity of the heart, based on periodic filling and emptying two rubber blowers a radioactive solution by means of two pistons in cylinders operated by two specially shaped cams. The phantom enables to control the correctness of the method and a computer programme for evaluating parameters of heart activity (ejection fraction, end-diastolic volume, pulse volume etc.) in balanced ventriculography performed by a scintillation camera coupled with a computer. The performed measurements have shown a very good agreement of the evaluated parameters with those set at the phantom. The designed device is of an original construction and was granted the author's certificate No. 238,298.
More Related Videos
06:48Three-Dimensional Echocardiographic Method for the Visualization and Assessment of Specific Parameters of the Pulmonary Veins
Published on: October 28, 2020
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021