An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
Ex Vivo HIFU Experiments Using a $32 \times 32$ -Element CMUT Array.
This study introduces a 32x32-element 2D capacitive micromachined ultrasonic transducer (CMUT) array for high-intensity focused ultrasound (HIFU) therapy. The developed CMUT array successfully created lesions in ex vivo tissues, demonstrating its therapeutic potential.
Area of Science:
- Biomedical Engineering
- Ultrasound Technology
- Medical Devices
Background:
- High-intensity focused ultrasound (HIFU) offers noninvasive therapeutic treatment for various diseases.
- Capacitive micromachined ultrasonic transducers (CMUTs) present advantages over traditional piezoelectric transducers for HIFU applications.
- Developing advanced transducer arrays is crucial for enhancing HIFU efficacy and precision.
Purpose of the Study:
- To design and fabricate a 32x32-element 2D CMUT array for HIFU applications.
- To investigate a novel element grouping strategy to simplify system complexity for addressing 1024 transducer elements.
- To evaluate the performance of the CMUT array in generating therapeutic ultrasound pressures and creating lesions in biological tissues.
Main Methods:
- Fabrication of an 8x8 mm^2, 32x32-element 2D CMUT array.
- Implementation of an element grouping scheme into eight HIFU channels based on phase delay for simplified addressing.
- Utilizing a custom application-specific integrated circuit (ASIC) for channel control.
- Characterization of transducer performance, including surface pressure and focal pressure measurements.
- Ex vivo tissue experiments to demonstrate lesion creation capabilities.
Main Results:
- The CMUT array was designed to focus at an 8-mm depth with a 5-MHz excitation frequency.
- A surface pressure of 1.2 MPa (peak-to-peak) was achieved with 40-V DC bias and 60-V peak-to-peak AC excitation.
- The focal peak-to-peak output pressure reached 8.5 MPa, sufficient for lesion generation within a minute based on thermal simulations.
- Successful ex vivo lesion creation was demonstrated in bovine muscle and liver tissue.
Conclusions:
- The developed 32x32-element 2D CMUT array is suitable for HIFU therapeutic applications.
- The proposed element grouping strategy effectively reduces system complexity for addressing a large number of elements.
- The CMUT array demonstrates significant potential for noninvasive tissue ablation and lesion generation.