Validation of continuous-wave Doppler measurements of mitral valve gradients during exercise--a simultaneous
W Voelker1, R Jacksch, H Dittmann
1Department of Cardiology, Tuebingen University, F.R.G.
Abstract:
In patients with mitral stenosis, continuous-wave Doppler measurements of the maximal transmitral inflow velocity can be converted into transvalvular pressure gradients using the modified Bernoulli equation. Because of close correlations between Doppler- and catheter-measured gradients this method has become a valuable tool in non-invasive evaluation of mitral stenosis at rest. However, in some patients, exercise studies are necessary to determine the haemodynamic significance of the valve stenosis. The accuracy of continuous-wave Doppler in this setting has not yet been validated. Thus, in 20 selected patients with pure or predominant mitral stenosis, continuous-wave Doppler echocardiography was performed during left- and right-heart catheterization. At rest and during submaximal bicycle exercise, Doppler and pressure measurements were simultaneously performed. The Doppler gradient was calculated according to the modified Bernoulli equation while the mean manometric gradient was determined from the simultaneous pulmonary wedge and left ventricular pressure curves. Exercise caused a significant increase of cardiac output (4.8 +/- 1.3 to 5.7 +/- 1.31 min-1) and heart rate (59.7 +/- 12.0 to 95.3 +/- 14.3 beats min-1). The mean Doppler gradient increased from 6.8 +/- 2.1 to 12.2 +/- 3.2 mmHg. The manometric gradient showed a comparable increase of 9.5 +/- 2.4 to 17.2 +/- 3.7 mmHg, respectively. Correlation between Doppler and manometric data was close (y = 0.79x-0.67,r = 0.90,SEE = 0.97 mmHg) at rest and still good during exercise (y = 0.71x-0.10,r = 0.82,SEE = 1.97 mmHg). Thus, in some patients with borderline resting gradients and valve areas, exercise Doppler might allow further identification of the haemodynamic severity of mitral stenosis.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment: