Extent of agreement between the body fluid model of Sysmex XN-20 and the manual microscopy method
Wei-Hua Huang1, Lin-Peng Lu2, Kang Wu1
1Department of Laboratory Medicine, Changhai Hospital, The Second Military Medical University, Shanghai, China.
Background:
Although the correlations concerning cellular component analysis between the Sysmex XN-20 body fluid (BF) model and manual microscopy have been investigated by several studies, the extent of agreement between these two methods has not been investigated.
Methods:
A total of 90 BF samples were prospectively collected and analyzed using the Sysmex XN-20 BF model and microscopy. The extent of agreement between these two methods was evaluated using the Bland-Altman approach. Receiver operating characteristic (ROC) curve analysis was employed to evaluate the diagnostic accuracy of high-fluorescence (HF) BF cells for malignant diseases.
Results:
The agreements of white blood cell (WBC), red blood cell (RBC), and percentages of neutrophils, lymphocytes, and monocytes between the Sysmex XN-20 BF model and manual microscopy were imperfect. The areas under the ROC curves for absolute and relative HF cells were 0.67 (95% confidence interval [CI]: 0.56-0.78) and 0.60 (95% CI: 0.48-0.72), respectively.
Conclusion:
Due to the Sysmex XN-20 BF model's imperfect agreement with manual microscopy and its weak diagnostic accuracy for malignant diseases, the current evidence does not support replacing manual microscopy with this model in clinical practice.