Cumulative Radiation Exposure With EOS Imaging Compared With Standard Spine Radiographs
T David Luo1, Anthony A Stans1, Beth A Schueler2
1Department of Orthopedic Surgery, Mayo Clinic, 200 1st Street SW, Rochester, MN 55905, USA.
Study Design:
Retrospective comparative study.
Objectives:
This study sought to estimate the total radiation exposure to scoliosis patients during the entire treatment course using standard imaging techniques versus EOS posteroanterior (PA) and anteroposterior (AP) views.
Summary Of Background Data:
EOS is a slot-scanning X-ray system designed to reduce radiation exposure in orthopedic imaging. There are few independent studies comparing organ and total effective radiation dose from standard EOS PA, AP, and lateral imaging versus conventional projection radiographs for children with spinal deformity.
Methods:
A total of 42 skeletally immature idiopathic scoliosis patients were treated with bracing (21) or spinal fusion (21) and were followed to skeletal maturity. The number of scoliosis radiographs (PA and lateral) for each patient was recorded. A computerized dosing model was used to calculate estimated patient and organ doses for PA and lateral scoliosis X-rays taken with EOS or computed radiography with a filter (CR) or without a filter (CRF). Assuming that each X-ray taken delivered the same radiation as the phantom calculation, the authors estimated the total effective and organ dose that each adolescent would have received using EOS, CR, or CRF. Annual background radiation is 3 mSv.
Results:
Mean number of radiographs per patient was 20.9 (range, 8-43). Patients who underwent surgical treatment had a significantly greater number of X-rays than those who were braced (27.3 vs. 14.5; p < .001). Assuming all films were CR, the mean cumulative dose was estimated at 5.38 mSv. With standard EOS films, the mean cumulative estimated dose was 2.66 mSv, a decrease of 50.6%. An AP versus PA EOS radiograph resulted in an 8 times higher radiation dose to the breasts and 4 times higher dose to the thyroid.
Conclusions:
The standard EOS imaging system moderately reduced the total radiation exposure to skeletally immature scoliosis patients. Over the entire treatment course, this represented 2.72 mSv mean reduction or 0.91 years of background radiation. Posteroanterior films significantly reduced breast and thyroid dose.
More Related Videos
09:41Imaging CD19+ B Cells in an Experimental Autoimmune Encephalomyelitis Mouse Model using Positron Emission Tomography
Published on: January 20, 2023
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
X-ray Imaging
Biological Effects of Radiation
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...