Biomechanical Assessment of Reduction Forces Measured During Scoliotic Instrumentation Using Two Different Screw
Mark Driscoll1, Jean-Marc Mac-Thiong1, Hubert Labelle1
1Spinologics, Inc., 289 Desaulniers, Quebec J4P 1M8, Canada.
Study Design:
Biomechanical finite element models simulated deformity correction using pedicle screw instrumentation and measured forces at the screw-vertebra interface.
Objectives:
Compare 2 different screw designs with respect to reaction forces at screw-vertebra interfaces during scoliosis correction maneuvers.
Summary Of Background Data:
Pedicle screw developments strive to enhance surgical techniques and improve patient safety. It is believed that a screw with increased lateral angulation and reduction tabs enables a more gradual correction, more effectively distributes corrective forces over multiple levels, and reduces forces at screw-vertebra interfaces compared with standard polyaxial screws.
Methods:
We selected 3 scoliotic patients and reconstructed their preoperative spinal profiles as finite element models using radiographic clinical measures. The osteoligamentous models were programmed and validated with mechanical properties from published literature. We used postoperative radiographs to determine instrumented levels and calibrate disc properties to corroborate simulated results with clinical data. We alternatively examined favored angle (FA) screws and polyaxial (PA) screws using correction steps characteristic to their design. We also explored sensitivity of screw forces consequent to misalignment with adjacent screws.
Results:
Simulated postoperative spinal profiles on average adhered to clinical measures within 5°. We observed no significant differences in simulated corrective profiles between screw types (5° or less). Compared with PA screws, FA screws reduced peak pullout and lateral forces by 27% and 35%, respectively, and correspondingly reduced mean pullout and lateral forces by 48% and 40%, respectively. Changes in peak and average pullout forces resulting from screw misalignment were 56% and 82% less, respectively, with FA screws.
Conclusions:
This analysis demonstrated reduced screw-vertebra peak and mean forces when using a pedicle screw with a favored angle bias and reduction tabs to correct scoliosis. Compared with PA screws, FA screws provide similar correction, decrease forces applied at the screw-vertebra interface, and are more forgiving if misaligned.
More Related Videos
09:07An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
05:57A Teleoperated Robotic System-Assisted Percutaneous Transiliac-Transsacral Screw Fixation Technique
Published on: January 6, 2023
Related Concept Videos
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Screw: Problem Solving
To evaluate the...