Devendra T Mourya1, Hamish C Shahani2, Pragya D Yadav1
1Maximum Containment Laboratory, Microbial Containment Complex, National Institute of Virology, Pune, India.
Background & Objectives:
Various conventional methods such as gaseous, vapour and misting systems, fogging, manual spray and wipe techniques employing a number of chemical agents are used for decontamination of enclosed spaces. Among all these methods, use of aerosolized formaldehyde is the most preferred method due to cost-effectiveness and practical aspects. However, being extremely corrosive in nature generating very irritating fumes and difficulty in maintaining a high level of gas concentration, many laboratories prefer the vaporization of hydrogen peroxide (H 2 O 2 ) as an alternative. We present here the results of using H 2 O 2 vapour in combination with plasma irradiation for quick decontamination of closed chambers.
Methods:
The present study describes a decontamination method, using plasma irradiation in combination with H 2 O 2 (5%). Effect of plasma irradiation and H 2 O 2 on the viability of bacterial spores (Bacillus subtilis), Chikungunya and Kyasanur Forest Disease viruses was assessed.
Results:
Data suggest that with the combination of H 2 O 2 vapour and plasma irradiation, within short time (three minutes), decontamination of surfaces and space volume could be achieved. Although it showed damage of spores present on the strips, it did not show any penetration power.
Interpretation & Conclusions:
The results were encouraging, and this method was found to be efficient for achieving surface sterilization in a short time. This application may be useful in laboratories and industries particularly, those working on clean facility concept following good laboratory and manufacturing practices.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview