Related Experiment Video
Updated: Mar 10, 2026

Diffuse Reflectance Infrared Spectroscopic Identification of Dispersant/Particle Bonding Mechanisms in Functional Inks
Published on: May 8, 2015
Portable near-infrared instruments: Application for quality control of polymorphs in pharmaceutical raw materials and
Vitor Hugo da Silva1, Jailson José da Silva2, Claudete Fernandes Pereira1
1Departamento de Química Fundamental, Universidade Federal de Pernambuco, Recife, PE, 50740-560, Brazil.
Abstract:
This work presents an evaluation of the analytical performance of three different portable near-infrared (NIR) instruments (denominated Port.1, Port.2 and Port.3) for quantifying mebendazole polymorphs (A, B and C) in pharmaceutical raw materials using multivariate calibration models. The performance of the portable instruments was compared with a benchtop one (FT-NIR Frontier spectrometer). In addition, calibration transfer between the benchtop and one of the portable instruments was also performed. For polymorph A, the Port.1 presented the lowest RMSEP value (1.01% w/w) even when compared to the FT-NIR instrument. For polymorphs B and C, the same Port.1 instrument presented RMSEP values of 2.09% w/w and 2.41% w/w, respectively, which were statistically similar to those obtained with the benchtop instrument. The LOD ranges (3.9-5.5 for polymorph A, 3.6-5.1 for polymorph B and 5.7-7.7 for polymorph C) obtained with the Port.1 was higher than those achieved with the benchtop NIR instrument, with high spectral resolution, signal-to-noise ratio and better wavelength reproducibility. Calibration transfer was performed between the benchtop NIR and Port.1 instruments. According to the results, the transferability of models is possible. The results obtained for complete recalibration of the portable instrument and those for the benchtop are comparable. The methods developed demonstrated a flexible, easy, cheap and fast way for quality control of MBZ polymorphs in incoming material, mainly in pharmaceutical laboratory chains.
More Related Videos
06:54Author Spotlight: Advances in Nanoscale Infrared Spectroscopy to Explore Multiphase Polymeric Systems
Published on: June 23, 2023
08:22Calibration-free In Vitro Quantification of Protein Homo-oligomerization Using Commercial Instrumentation and Free, Open Source Brightness Analysis Software
Published on: July 17, 2018
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Applications of IR Spectroscopy: Overview
IR Spectrometers
Flame Photometry: Lab
Infrared (IR) Spectroscopy: Overview
Different compounds display unique properties due to their...