Characterizing the reduction of stimulation artifact noise in a tripolar nerve cuff electrode by application of a
Parisa Sabetian1, Bita Sadeghlo2, Chengran Harvey Zhang2
1Institute of Biomaterials and Biomedical Engineering, University of Toronto, 164 College St, Room 407, M5S 3G9, Toronto, ON, Canada.
Abstract:
Tripolar nerve cuff electrodes have been widely used for measuring peripheral nerve activity. However, despite the high signal-to-noise ratio levels that can be achieved with this recording configuration, the clinical use of cuff electrodes in closed-loop controlled neuroprostheses remains limited. This is largely attributed to artifact noise signals that contaminate the recorded neural activity. In this study, we investigated the use of a conductive shield layer (CSL) as a means of reducing the artifact noise recorded by nerve cuff electrodes. Using both computational simulations and in vivo experiments, we found that the CSL can result in up to an 85% decrease in the recorded artifact signal. Both the electrical conductivity and the surface area of the CSL were identified as important design criteria. Although this study shows that the CSL can significantly reduce artifact noise in tripolar nerve cuff electrodes, long-term implant studies are needed to validate our findings.