Pre-operative Simulation of the Appropriate C-arm Position Using Computed Tomography Post-processing Software Reduces
E Stahlberg1, M Planert1, N Panagiotopoulos1
1Department for Radiology and Nuclear Medicine, University Hospital of Schleswig Holstein, Lübeck, Germany.
Objective/Background:
The aim was to evaluate the feasibility and efficacy of a new method for pre-operative calculation of an appropriate C-arm position for iliac bifurcation visualisation during endovascular aortic repair (EVAR) procedures by using three dimensional computed tomography angiography (CTA) post-processing software.
Methods:
Post-processing software was used to simulate C-arm angulations in two dimensions (oblique, cranial/caudal) for appropriate visualisation of distal landing zones at the iliac bifurcation during EVAR. Retrospectively, 27 consecutive EVAR patients (25 men, mean ± SD age 73 ± 7 years) were identified; one group of patients (NEW; n = 12 [23 iliac bifurcations]) was compared after implementation of the new method with a group of patients who received a historic method (OLD; n = 15 [23 iliac bifurcations]), treated with EVAR before the method was applied.
Results:
In the OLD group, a median of 2.0 (interquartile range [IQR] 1-3) digital subtraction angiography runs were needed per iliac bifurcation versus 1.0 (IQR 1-1) runs in the NEW group (p = .007). The median dose area products per iliac bifurcation were 11951 mGy*cm2 (IQR 7308-16663 mGy*cm2) for the NEW, and 39394 mGy*cm2 (IQR 19066-53702 mGy*cm2) for the OLD group, respectively (p = .001). The median volume of contrast per iliac bifurcation was 13.0 mL (IQR: 13-13 mL) in the NEW and 26 mL (IQR 13-39 mL) in the OLD group (p = .007).
Conclusion:
Pre-operative simulation of the appropriate C-arm angulation in two dimensions using dedicated computed tomography angiography post-processing software is feasible and significantly reduces radiation and contrast medium exposure.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...